TT-4500 Low Speed Centrifuge

Category: .

Product Description

TT-4500 Low Speed Centrifuge is a contemporary low speed benchtop centrifuge designed for operation with centrifuge tubes and conical tubes up to 50ml. It is intended for biological sample separation with adjustable speed range of 300-4500rpm which is widely used in various profile laboratories including biomedical, pharmaceutical, microbiology and molecular biology.

Three autoclavable angle rotor options are available to accommodate common tube sizes (5ml, 7ml, 10ml, 15ml and 50ml). The device enables rapid and gradient acceleration (30 sec) and deceleration (20 sec), thus eliminating time-consuming and centrifugation process.

You may enjoy a peace of mind while operating the unit as it comes with safety features such as double-layered safety lid with auto-lock function, alarm and stop automatically when over-speed, over-heat and imbalance occurred. Operating noise level is also low at ≤56 dB to avoid any unwanted distraction in the laboratory.

Features

  • Maximum speed: 4500 rpm (2,490 x g)
  • Intuitive LCD control panel
  • Multiple time setting modes: Short mix, digital time setting and continuous
  • Rapid and gradient acceleration and deceleration
  • Automatic lid release upon program completion
  • Multiple safety features: double-layered safety lid, auto-lock function, audio alert and automatically stop when over-speed, over-heat and imbalance occurred
  • Quiet operation for peaceful working environment
  • Run-to-run reproducibility, timer starts when the selected rotational speed is reached
  • Fully autoclavable rotors (120°C)
  • Rust-free and easy to clean stainless steel rotor chamber

Application

  • Biological sample separations of cellular materials, lipid cell, stem cell, blood, urine etc.
  • Cell harvest from cultures (protein expression, bioprocessing)
  • Pelleting for bacteria, animal and plant cells as well as membrane organelles or fractions
  • Biomedical, clinical, pharmaceutical, proteomics, biochemistry, microbiology, molecular biology, chemical laboratories

Technical Data

SpecificationsTT-4500
Low Speed Centrifuge
Speed300 - 4500 rpm (100rpm steps)
Maximum RCF2490 × g (increment: 100 × g)
RPM/RCF conversionYes
Speed accuracy±20rpm
Rotor capacity5 ml/ 7 ml x 16; 10/ 15 ml x 12; 50 ml x 6
Digital time setting1sec - 99min 59sec or continuous
Short spinYes
Automatic lid lockYes
Automatic lid release after completionYes
Alarm and automatic stopYes, during over-speed, over-heat and imbalance
Noise level≤56dB
Acceleration time to max. speed30sec
Braking time from max. speed20sec
Soft acceleration/decelerationYes
Motor typeBrushless DC motor
VoltageAC 110-240V, 50/60Hz
Power120W
Dimension (W x D x H)300 x 431 x 178 mm
Net Weight17.0 kg
Permissible ambient temperature4 - 40 oC
Permissible relative humidity≤70%

Resources

Documents

Data Sheet : TT-4500 Low Speed Centrifuge

Video

TT-4500 Low Speed Centrifuge

Publication

Noutsa, B.S. et al. (2022) Chemical Characterization and Antibacterial Properties of Fontitrygon margarit (Gunther,1870) Liver Oil. Journal of Lipids. Volume 2022, Article ID9369387, 10 pages

Ordering Info

Cat. No.TT-4500 Low Speed Centrifuge
TT-4500Low Speed Centrifuge TT-4500, 300-4500 rpm, AC 110-240V

Optional Accessories

Cat. No.Rotor for Low Speed Centrifuge
TT-4500-R01Angle rotor for 5 ml/ 7 ml tubes x 16
TT-4500-R02Angle rotor for 10 ml/ 15 ml tubes x 12
TT-4500-R03Angle rotor for 50 ml tube x 6

FAQ

1)  How to open the lid in case of power failure?
Upon power failure while running, the lid will have to be opened manually. Wait until it has come to a full stop. Then, unplug the power cord. Press the shaft on the left hole on the equipment and the lid will open. Please refer to the manual for more information.

2)  What is the minimum speed increment?
The speed can be increased in increments of 100 rpm.

3) What are the maximum values for acceleration and deceleration times?
The acceleration time to reach the maximum rotational speed is approximately 30 sec while the deceleration time from the maximum rotational speed is about 20 sec.

4) How do we maintain the rotors and interior of the centrifuge?
The rotors and interior should be cleaned on a monthly basis. Unplug the power cord and remove the rotor. Use a soft, moist cloth and neutral cleaning solutions (alcohol or alcohol-based disinfectant) to clean the rotor and interior chamber. For disinfection purpose, diluted 70% ethanol can be used. Do not pour water directly into the chamber. Any spill should be wiped immediately before using. After wiping or rinsing the rotor with clean water, leave it to dry before installing.

5) What kind of solution should we avoid to clean the centrifuge?
Do not use detergents that contain chlorite ions as it often contributes to corrosion. Avoid using acetone and caustic detergents as well as dishwashers because some aggressive cleaning agents may cause corrosion.

6) Are the rotors autoclavable?
The rotors are fully autoclavable within 120°C.

7) How do I balance my samples?
Always balance your samples because running a centrifuge with unbalanced load may lead to permanent damage of the centrifuge. Also check the rotor holes for any spill or leftover residue from previous run, remove if any, as they may lead to unbalanced operation. If you only have one sample to run, make sure to put another tube with the same volume of water in the opposite hole. Balancing by mass is recommended if the density of sample and water is different. This practice is critical and may just save your life and others in the lab.

8) What other tips or safety precautions should we take while using the centrifuge?
Always place the centrifuge on a flat and firm surface. Never try to open the lid while it is running. If the centrifuge starts wobbling or shaking drastically, pull off the power cord immediately. Always check the tubes for cracks because cracks can cause spillage while running and lead to imbalance operation.