11 Feb
2019
Product Discontinuance Notification ND-2800-OD NanoDOT Nucleic Acid Analyzer
Dear valued customers,
We would like to announce that Hercuvan Lab Systems Sdn Bhd has decided to discontinue supply of ND-2800-OD NanoDOT Nucleic Acid Analyzer effective from 10th January 2019.
The alternative supply of the current ND-2800-OD NanoDOT Nucleic Acid Analyzer will be replaced by the new ND-2800- ODJ NanoDOT Nucleic Acid Analyzer.
Please find the following tables for the scope of changes in ordering information and technical specification.
| Current Model | New Model | |
| Catalogue number | ND-2800-OD | ND-2800-ODJ |
| Sample size | 0.5 μl – 2μl (2μl advised) | 1 μl – 2 μl (2μl advised) |
| Path length | 0.2 mm (high concentration) 1.0 mm (ordinary concentration) | 0.5 mm |
| Light source | Xenon flash lamp | UV LED |
| Detector type | Silicon photodiode | UV silicon photocell |
| Wavelength range | 230nm, 260nm, 280nm | 260nm, 280nm |
| Wavelength accuracy | ±6 nm | – |
| Absorbance range (10 mm equivalent absorbance) | 0.01 – 80 A | 0.02 – 80 A |
| Detection range | 10 – 4,000 ng/μl (dsDNA) | 10 – 2,500 ng/μl (dsDNA) |
| Detection time | <10s | 10 – 2,500 ng/μl (dsDNA) |
New Ordering Information:
| Cat. No. | Nucleic Acid Analyzer |
| ND-2800-ODJ | Nano DOT Nucleic Acid Analyzer ND-2800-ODJ, DC 24V |
The last date of order acceptance for current ND-2800-OD is on 17th January 2019. The order shall not be fulfilled if there is no inventory held by the Company.
